Installation
Page 7
Manual BÜHLER BU4010ff
Device designation
4010
Housing
Stainless steel with 40 mm insulation,
(material 1.4301/ SS304) / PS / PC (GF10)
Option:
(material 1.4571/ SS316Ti; SS304 EPOXY-coated;
SS316Ti EPOXY-coated)
Thermostatic control
Self-contained cooling / heating with 4 settings, no-frost.
Temperature in sample compartment: 4°C (setting range: 0,0-9,9°C)
Control
Microprocessor control, foil keyboard, graphical display (128*64 Pixel),
back lit
Data memory
Non-volatile data memory: storage of sampling and malfunction data
like sample extractions, bottle changes, messages, external signals
Programming
12 user programs (can be edited freely)
Program start options
Immediately, at a certain time, by an external signal
Program stop options
End of sampling program after one program run, continuous operation
or x-runs
Pause mode
Interruption of program at any time
Overfilling protection
Adjustable from 1–999 samples/bottle
Interval setting
1 min. to 99 h 59 min. in steps of 1 minute
Pulse setting
1 to 9999 pulses/sample
Manual sample extraction
Possible at any time without interrupting the current program run
Program protection
Up to 5 years after voltage loss
Interface
Mini-USB, RS422/485, RS 232
Communication
Optional: TCP/IP RJ45 WEB board, 4-32GB SD/SDHC memorycard
Optional: Modbus or DP PROFIBUS connection
Languages
Multi-language, selectable
Signal inputs
• 2x analogue: 0/4-20 mA,
• 8x digital (flow, event, 1x free programmable)
Optional: expansion 4x digital, 3 free programmable and
8x Analogue Inputs 0-20mA / 0-10 V
Pulse-length min. 60 ms u. switch-level 7-24V, max. working
restistance 500 Ohm, signal-length max. 30 m
Signal outputs / status
messages
Max. 8 digital outputs, 1x collective message
Optional: expansion 8x digital, 5 are free programmable
Sampling method
-Vacuum system 20-350 ml, plastic vessel
-Vacuum system 20-500 ml, optional
-Vacuum flow-proportional 5-350 ml, optional
-Bypass system 20-250 ml optional
Single sample volume
accuracy
< 2,8 % at standard vacuum system
Suction height
Max. 8 m (at 1013 hPa and stagnant medium)
Pumping speed
>0,5 m/s at a suction height up to at least 7,8 m (at 1013 hPa); pump
capacity can be adjusted electronically.
Suction hose
PVC, L=7,5 m, ID=12 mm (ID 16 mm at flow-proportional option)
Max. hose length 30 m
Sampling modes
Time-related, flow-dependent, flow-proportional (option), event-related,
manual sample extraction
Bottle variants
Plastic
1 x 25 L, 1 x 50 L, 2 x 10 L
4 x 6,0 L, 4 x 10 L, 4 x 14 L, 4 x 20 L, 4 x 25 L
12 x 2,9 L, 24 x 1,0 L, 24 x 2,9 L
Glass
12 x 2,0 L, 24 x 1,0 L, 24 x 2 L
Overall dimensions
(Hxwxd)
1.290 (1.890*)x 690 x 645 mm
for bottle variants: 1x25 L, 1x50 L, 2x10 L, 2x22 L, 4x6 L, 4x10 L,
4x14 L, 12x2 L, 12x2,9 L, 24x1 L
1400 (2015*) x 930 x 850 mm
for bottle variants: 4x20 L, 4x25 L, 24x2,9 L, 24x2 L, 36x1 L