Troubleshooting – flow-proportional sampling system
If a fault develops in your sampler, we can only provide prompt assistance if you give us the type and
serial number of the apparatus concerned. You will find these data on the type plate.
The more precise your description of the fault is, the better our fault diagnosis will be.
Symptom
Possible cause
Action
Level electrodes do not react,
system is flooded
Adjustment of conductivity is
too low. Distilled water is used
as sample medium
Contact service department
No contact to electrodes
Check cable connections
Sampler does not extract
samples
System is leaking
Check whether all hoses and
connections are tight
Pump/diaphragm defective
Check pressure/vacuum of
pump
Metering vessel not tight, no
vacuum
Check whether metering vessel
is tight (union nut)
Level electrodes are bridged
Clean electrodes
Valve system/pinch valve does
not work correctly
Check valve system/pinch valve
or contact service dept.
Pump does not produce
pressure/vacuum
Diaphragm defective
Replace diaphragm
Hoses are kinked
Check hoses
Inaccurate sample volume
System has been initialized or
hardware has been changed
Calibrate