Nitratax plus sc, Anise sc/aise sc/nise sc, Nitratax plus sc anise sc/aise sc/nise sc – Hach-Lange PROGNOSYS User Manual
Page 15: Anise sc/aise sc/nise sc

NITRATAX plus sc
Message
Possible cause
Solution
Sensor,
Application
check
The signal level is too low. The sensor
possibly has a dirty window. The selected
path length is too long (if applicable). The
solid concentration or the turbidity in the
measurement medium can be too high.
• Clean sensor window fully.
• Examine the wiper.
• Examine the measuring medium with a
diluted sample.
• Contact technical support.
R meas ext The measurement absorbance is too high. • Clean sensor window fully. concentration. If the EM/ER values are • Contact technical support. ref ext The path length is too long. The wiper does Wiper blocked The wiper is mechanically blocked. The • Examine and clean the measurement window fully. • Complete a wiper test and do a stop position test. • Contact technical support. Wiper position Moist Moisture in the sensor is over limit: the Contact technical support immediately to R too high The automatic zeroing is not correct. Contact technical support for a zero point Replace The time for the wiper axle gasket cycles is Contact technical support to replace for a wiper Replace profile The time for wiper profile cycles is expired. • Replace the wiper profile. is completed. Motor cycle The time for motor cycles is expired. Contact technical support to replace the wiper Flash lamp The time for flashes is expired. Contact technical support to replace the flash Replace seals The annual change of the sensor enclosure Contact technical support to replace the sensor Error Collective error message Examine the error message on the controller. Warning Collective warning message ANISE sc/AISE sc/NISE sc Message 1 Possible cause Solution Instrument error The instrument caused an Examine the error message on the controller. Refer to Instrument warning The instrument caused a English 13
The measurement range is too high because
the signal level is too low. The sensor
possibly has a dirty window. Interferences
can be caused by other substances.
• Examine the wiper.
• Examine the solid content of sample.
• Use a cuvette test to examine the nitrate
>2.74, use a smaller path length.
not operate correctly. The absorbance is too
high because of nitrate concentration or other
UV-absorbing materials.
measurement window is not clean anymore.
No sample is pulled in the instrument. The
wiper is possibly in front of the window.
unknown
desiccant is expired. There is possibly a
problem with the gasket set for the sensor
sleeve or the wiper axle.
replace the gasket, so that the instrument
operates correctly.
calibration.
shaftseals
expired.
axle gasket.
• Set the counter when the maintenance task
motor.
replace
lamp.
gasket is necessary.
enclosure gasket.
Refer to the sensor documentation for
solutions.
error message.
the sensor documentation for solutions.
warning message.