Configure the sensor, Service and measurement value indicator messages, Amtax sc and phosphax sc – Hach-Lange PROGNOSYS User Manual
Page 13: Refer to

Configure the sensor
Use the PROGNOSYS menu to monitor specific settings or to change the blink mode.
1. Push MAIN MENU>RTC MODULES/PROGNOSYS>PROGNOSYS.
2. Select the applicable sensor.
3. Select an option.
Option
Description
MEAS. INDICATOR
Shows the measurement indicator bar in %.
DETAILS
Shows the message list for the measurement indicator. Example: R
shows the absorbance value in %.
SERVICE INDICATOR Shows the number of days that remain until a maintenance task must be completed.
DETAILS
Shows the message list for the service indicator. Example: replace the wiper profile:
1 day or replace the seals: 42 days
DEVICE
Shows the sensor name.
LOCATION
Shows the location name where the sensor is used.
PROGNOSYS VERS.
Shows the PROGNOSYS software version.
BLINK MODE MEAS< Enters the measurement indicator value in %. The measurement indicator bar
flashes when the value decreases below the given value. Make sure to enter a
higher value when the measurement is related to a control system or a sensitive
parameter. Range: 0 to 100% (default = 0%)
BLINK MODE SERV<
Enters individual number of days for a maintenance task. The service indicator bar
flashes when the value decreases below the given value. Range: 0 to 200 days
(default = 0 days)
Service and measurement value indicator messages
W A R N I N G
For safe use of the instrument, obey the precautions and the instructions in the sensor documentation.
This list gives a general overview for the service and measurement value indicator messages. Refer
to the sensor documentation for a more detailed list of maintenance messages.
AMTAX sc and PHOSPHAX sc
Message
Possible cause
Solution
Instrument error
The instrument caused an error
message.
Examine the error message on the controller. Refer to
the sensor documentation for solutions. Make sure that
the error is shown in the service menu and push
START.
Instrument
warning
The instrument caused a warning
message.
Warm-up phase
The inside of the instrument is too
cold (e.g., the enclosure door was
open at low external
temperatures).
Wait until the warm-up phase is completed. If the start-
up temperature was higher, the warm-up phase can
vary from a few minutes to an hour.
Cooling down
The instrument is overheated and
is cooling-down. If the instrument
version has a filter probe, the
compressor is deactivated in this
phase.
• Wait until the cooling-down phase is completed.
• Make sure that no blockage of the vent is caused.
• Clean or change the air filter.
• Complete a function test on the fan.
• Specify the correct operating temperature.
English 11