Standard operations – Hach-Lange DR2800 eco User Manual
Page 61

61
Standard Operations
8. Press OK. The link is established automatically and the
software is updated.
9. Press OK to return to the System Checks menu.
Note: When the instrument software has been updated, a prompt to restart
the instrument is displayed.
5.5.8 Program an own barcode-readable application with LCW 906
LCW 906 empty cuvettes can be used to program a
barcode-readable applications.
Several user applications can be assigned to this program number.
Note: Use the program name to assign the barcode application to
LCW 906.
1. Press Stored Programs in the Main Menu. Press Program
Options and then New.
2. Use the alphanumeric keypad to enter the Program Number
906.
3. A data overview is displayed, including the specifications of the
test that is to be revised.
4. Highlight the corresponding line in the data overview and press
Edit.
For general advices on updating program data, see
5.5.8.1 Evaluation of barcode-readable application with LCW 906
1. The instrument needs to display the Main Menu or the
Barcode Programs menu.
2. Insert the cuvette LCW 906.
3. Select the required application 906 from the displayed list and
press Start.
Note: Press Start Permanent if this application 906 is to apply to all
subsequent measurements.
To disable the function Start permanent:
Insert the cuvette LCW 906. Press Options>More>Permanent: On.