beautypg.com

3 lamp replacement – Hach-Lange DR2800 eco User Manual

Page 118

background image

118

Maintenance

7.3 Lamp replacement

CAUTION
To avoid a possible electric shock, disconnect the instrument
from the power source before servicing the lamp.

1. Remove the cuvette from the cell compartment.

2. Switch the instrument off.

3. Unplug the power cord.

WARNING
Burn Hazard. Wait until the lamp cools down. Contact with the
hot lamp can cause burns.

4. Turn the instrument over carefully and place it on a soft surface.

5. Use a screwdriver or a coin to remove the screw from the

cover.

6. Remove the cover and place it to one side.

7. If there is a battery in the instrument, push it out and put it to

one side too, see

section 7.2.1 on page 115

.

8. Use a screwdriver (slotted or cross-headed) to remove the

screws from the lamp fitting.