2 principle of operation, 4 items supplied, 1 identification of the operating instructions – Hach-Lange Filter probe sc User Manual
Page 10

8
General information
2.3.1.1 Rating plate for the filtration probe sc
You will find the rating plate at the top end of the supply cable. On 
the rating plate you will find
•
The model number and the serial number,
•
Information on the voltage and frequency for the electrical 
connection and
•
Information on the brief peak power consumption.
2.3.2 Principle of operation
The filtration probe sc draws in liquid through a filter system and 
then pumps this sample flow to an analysis instrument connected in 
series. After the analysis of the sample, the analysis instrument 
pumps the sample (possibly mixed with chemicals) back to the tank 
sampled or to a drain. 
A cleaning device ensures the filter modules remain porous. The 
filtration probe sc has a compressed air tube that terminates 
underneath the filter modules. When compressed air flows out of 
the air openings underneath the filter modules, the rising air 
bubbles remove soiling from the filter modules. In this way the filter 
modules remain porous and manual cleaning is largely 
unnecessary.
2.4 Items supplied
Standard items supplied:
•
Filtration probe sc
•
These operating instructions
•
2 filter modules
•
Small parts, cable, tubing and connector set and
•
One cleaning sponge
2.4.1 Identification of the operating instructions
You will find a DOC number on the cover sheet so that you can 
re-order these operating instructions. You can identify the issue of 
the operating instructions using the date in the DOC number.