7. If the display shows Check Standard Passed, the check standard measurement is
within the accepted limits set by the administrative user. Select Done to continue with
the sample measurement.
8. If the display shows Check Standard Failed, the measurement is outside of
accepted limits set by the administrative user and a recalibration is recommended. If
the acceptance criteria is set to Cal Expires on Failure: Yes, the display shows the
calibration icon and a question mark until the probe is recalibrated. To correct the
probe calibration and status indicator, calibrate the probe (refer to
on page 3).
Advanced operation
Parameter-specific settings can be changed through the Full Access Options menu.
Details about menu navigation, available options and how to change them are given in
the screens, tables and procedures throughout this section.
The settings that can be changed are shown in
Table 1 Parameter-specific settings
Setting
Options
Measurement Options
•
Resolution
•
Upper and lower range limits
Calibration Options
•
Buffer set
•
Buffer set values (if Custom Buffer Set selected)
•
Reminder
•
Minimum Cal points
•
Slope limit
Check Standard Options
•
Standard (temperature compensated buffer or custom at 25 °C)
•
Reminder
•
Acceptance criteria
•
Standard value at 25 °C (if custom standard selected)
Units
•
pH
•
mV
Change measurement options
Methods are groups of default or user-defined settings relevant to specific applications. If
the meter is set to the default method and the Modify Current Settings option is chosen, a
prompt for a new name is shown after the changes are entered. The settings are saved
with this name to distinguish them from the default method settings, which cannot be
changed. A saved method can be used instead of multiple adjustments to the individual
settings. Changes made to a user-defined method are automatically saved with the
existing name. Multiple methods can be saved for the same probe on each meter.
1. Make sure a probe is connected to the meter.
2. Push and select PHC101 Settings.