Hach-Lange PHC10115 User Manual User Manual
User manual, Safety information, Specifications

User Manual
DOC022.5
2.80023
Gel-filled pH Probe: Model PHC10101, PHC10103, PHC10105,
PHC10110, PHC10115 or PHC10130
Safety information
Precautionary labels
Read all labels and tags attached to the instrument. Personal injury or damage to the
instrument could occur if not observed. A symbol on the instrument is referenced in the
manual with a precautionary statement.
Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European
electrical equipment users must now return old or end-of-life equipment to the Producer for disposal at no charge to
the user.
Note: For return for recycling, please contact the equipment producer or supplier for instructions on how to return end-of-life
equipment, producer-supplied electrical accessories, and all auxiliary items for proper disposal.
Specifications
Note: Specifications are subject to change without notice.
Specifications
Details
Probe type
Digital combination non-refillable, gel-filled probe with double junction
reference and built-in temperature sensor
pH range
pH 2 to 14
pH resolution
User-selectable stabilization time and resolution—Fast: 0.1, Fast: 0.01,
Medium: 0.01, Slow: 0.01 or Slow: 0.001
Slope
-59 mV/pH (90 to 110% at 25 °C (77 °F) per Nernstian theoretical value)
Operating temperature range
0 to 50 °C (32 to 122 °F)
Storage temperature range
5 to 40 °C (41 to 104 °F)
Junction
Open
Reference type
Ag/AgCl
Sodium (alkalinity) error
-0.6 pH at pH 12.6 in 1 M NaOH
Temperature accuracy
±0.3 °C (±0.54 °F)
Minimum sample depth
20 mm (0.79 in.)
Dimensions (standard)
Diameter: 12 mm (0.47 in.)
Length: 175 mm (6.89 in.)
Cable length: 1 or 3 m (3.28 or 9.84 ft)
Dimensions (rugged)
Diameter: 46 mm (1.81 in.)
Length: 223 mm (8.73 in.)
Cable length: 5, 10, 15 or 30 m (16.40, 32.81, 49.21 or 98.42 ft)
Cable connection
M12 digital output and connector compatible with HQd meters
Product overview
The PHC101 series probe is a non-refillable, gel-filled combination pH probe with a built-
in temperature sensor (
). The standard PHC10101 or PHC10103 probe comes
with a 1 or 3 m (3.28 or 9.84 ft) cable and is intended for laboratory use. The rugged
PHC10105, PHC10110, PHC10115 or PHC10130 comes with a 5, 10, 15 or 30 m (16.40,
1