Hach-Lange PHC10115 User Manual User Manual
Page 11

Message or symptom
Possible cause
Action
Connect a probe or probe
requires service
Probe not connected properly
Disconnect, then connect the probe. Tighten the locking
nut.
Software not updated
To download the most current version of the software,
refer to the applicable product page on the manufacturer's
website.
Refer to the HQd Series meter manual.
Large number of methods stored
Continue to let probe connect. Do not disconnect the
probe.
Damaged probe
Make sure there is connectivity with another probe or
meter to confirm isolated issue with probe. Contact a
Technical Support Representative.
Standard not recognized
error
Tape not removed from the
reference junctions
Remove the tape.
Standard not recognized
error
Probe storage cap or soaker
bottle not removed.
Remove the probe storage cap or soaker bottle.
Incorrect or contaminated buffer
solution
Use fresh buffer solution as specified in the method.
pH reading and/or mV
reading is same for all
solutions
Tape not removed from the
reference junctions
Remove the tape.
pH reading and/or mV
reading is same for all
solutions
Storage cap or soaker bottle not
removed
Remove the storage cap or soaker bottle.
Electrical issue
Contact a Technical Support Representative.
Slow stabilization time
Tape not removed from the
reference junctions
Remove the tape.
Slow stabilization time
Contaminated glass sensor
on page 9).
Poor contact between reference
junction and solution
Shake the probe in the solution from side to side to refresh
the reference junction.
Probe not prepared for sample
For the best stabilization time, soak the probe in the
sample for 10 to 15 minutes before doing a sample
measurement.
Low sample temperature or
temperature difference between
samples
Check the sample temperature. The lower the
temperature or the greater the difference of temperatures
between samples, the longer the stabilization time will be.
Air bubbles around inner
reference electrode
Gently tap the probe with hand or shake the probe
downward to remove any air bubbles.
Calibration errors
Calibration not done correctly
Recalibrate using freshly prepared pH buffers.
Contaminated glass sensor
on page 9).
Slope exceeds the criteria for %
of theoretical (as defined in the
method slope limit)
1.
Widen the slope limits by changing the
PHC101 calibration settings and method.
2.
Recalibrate the probe.
3.
Run a check standard to check the performance of
the probe.
11