Inputs/outputs menu – Hach-Lange ORBISPHERE Model 51x User Manual User Manual
Page 76

74
Inputs/Outputs Menu
Analog output characteristics (continued)
“Tri-linear” analog 
output
The "Tri-linear" output brings benefits over the “Linear output” discussed 
before. It is illustrated below (4-20 mA output is shown, 0-20 mA or 0-5 V 
settings are similar).
Compared to the “Linear” mode, the expected range of measure is Range 2. 
A Range 1 and 3 are available to show the measures falling out of this 
Range 2, but normally at a lower resolution. Expected measurements for the 
measured process are supposed to be in Range 2 most of the time, and in 
Range 1 or 3 occasionally (problems, calibration, line stop, etc.). The 
benefits are:
•
The PLC can compute the measurement over a large range (1, 2 and 3).
•
The PLC can compute a higher resolution signal for the expected 
measuring range (Range 2: MH > M > ML). 
•
Carefully selecting the set points allows for an individual resolution for 
each range, so a different resolution can be applied to Range 1, 2 and 3, 
allowing to tailor the analog output to the actual conditions.
As before, the downside is that any measure below or over the Range 1, 2 
and 3 will have the same signal locked at 4 mA and 20 mA respectively, but 
Range 1, 2 and 3 should cover a larger range than in the “Linear” mode. 
Settings must made in balancing these aspects.
Settings
For each output, set MLL, ML, MH, and MHH in the current measuring unit 
(e.g. °C for a temperature output). When a compound unit is selected, the 
smallest unit will be used (e.g. ppb for a "ppm-ppb" compound unit). Also set 
AOL (Analog Output Low) and AOH (High) in mA (or Volts).
These points should be set keeping in balance the following conditions (see 
illustration above):
•
The smaller is Range 2, the better is the analog signal resolution within 
the expected range of measure.
•
Size of Range 1 and 3 should be set to deliver an adequate level of 
resolution for the measures falling out of the expected range of 
measure. 
•
In Range 0 the analog outputs only shows that measurement is below 
the MLL value. Similarly in Range 4 the analog output only shows that 
measurement is over the MHH value. 
The formula to compute the measurement knowing the current or the 
voltage and the resolution R is given in the following table:
