Specifications
Specifications are subject to change without notice.
pH/ORP sensors
8350.3
8350.4
8350.5
8351
Application
Measuring pH in
samples containing
hydrofluoric acid
Measuring pH at
high temperatures
Measuring pH in a
waste water
environment
Measuring ORP
Material
PPS
PPS
CPVC
PPS
Measurement range
0—12 pH
0—14 pH
0—12 pH
± 1500 mV
Maximum
temperature
110 °C
110 °C
80 °C
110 °C
Maximum pressure
10 bar
10 bar
10 bar
10 bar
Repeatability (week)
< 0.05 pH
< 0.05 pH
< 0.05 pH
< 0.05 pH
Offset @ pH7
± 0.34 pH
± 0.34 pH
± 0.34 pH
N/A
Slope
56—61 mV/pH
56—61 mV/pH
56—61 mV/pH
N/A
Reference
impedance @ 25 °C
< 50 Kohm
< 50 Kohm
< 50 Kohm
< 50 Kohm
Glass impedance @
25 °C
100—150 Mohm
150—500 Mohm
50—250 Mohm
N/A
8362
8417
Application
Measuring pH in pure or ultra pure
water
Measuring pH in industrial and
waste water treatment plants
Material
316L stainless steel
Glass membrane, ceramic junction
Measurement range
2—12 pH
0—14 pH
Maximum temperature
80 °C
110 °C
Maximum pressure
16 bar @ 25 °C;
6 bar @ 100 °C
10 bar @ 25 °C
Repeatability (24 hours)
< 0.01 pH
0.02 pH
Input impedance
> 10
6
Mohm
100 Mohm
Antimony electrode sensors
8346 / 8347
Application
Measuring pH or ORP in dirty water
Material
Polypropylene
Maximum temperature
80 °C (pH measurements with 8346)
120 °C (pH measurements with 8347)
50 °C (ORP measurements)
Sample pressure
Atmospheric
Maximum pressure (electrolyte container)
5 bar