Table 5 Warning list for pH and ORP sensors (continued)
Warning
Description
Resolution
PH TOO LOW
The measured pH is < 0
Calibrate or replace the sensor.
ORP TOO LOW
The measured ORP value is
< –2100 mV
OFFSET TOO HIGH
The offset is > 8 (pH) or 200 mV
(ORP)
Follow the maintenance procedures
for the sensor and then repeat the
calibration.
OFFSET TOO LOW
The offset is < 6 (pH) or –200 mV
(ORP)
SLOPE TOO HIGH
The slope is > 60 (pH)/1.3 (ORP)
Repeat the calibration with a fresh
buffer or sample.
SLOPE TOO LOW
The slope is < 54 (pH)/0.7 (ORP)
Clean the sensor, then repeat the
calibration.
TEMP TOO HIGH
The measured temperature is >
100 °C
Make sure that the sensor is
configured for the correct
temperature element.
TEMP TOO LOW
The measured temperature is <
0 °C
CAL OVERDUE
The Cal Reminder time has expired
Calibrate the sensor.
NOT CALIBRATED
The sensor has not been calibrated
Calibrate the sensor.
REPLACE SENSOR
The sensor has been in operation >
365 days
Replace the sensor.
CAL IN PROGRESS
A calibration was started but not
completed
Return to calibration.
OUTPUTS ON HOLD
During calibration, the outputs were
set to hold for a selected time.
The outputs will become active after
the selected time period.
FLASH FAILURE
External flash memory failed
Contact technical support.