Set up the discrete inputs – Hach-Lange POLYMETRON 9500 User Manual User Manual
Page 39

Figure 24 Frequency control function
1 High limit
4 Low limit
7 Selected source (y-axis)
2 Deadband
5 Cycle duration
3 Setpoint
6 Time (x-axis)
• Warning function
Option
Description
WARNING Sets the level for warning activation. Refer to the sensor manual for the numbers for individual
warning messages.
Set up the discrete inputs
Use these inputs to switch closure inputs or logic level voltage inputs.
1. Push the menu key and select Polymetron 9500 SETUP>DISCRETE INPUT SETUP.
2. Select the desired channel (INPUT 1, INPUT 2 or INPUT 3) and push enter.
3. Select a control logic option and push enter.
Option
Description
DISABLE This channel is disabled and not used.
ON/HIGH
This channel is active when either the switch input is On (or closed), or the logic level voltage
input is at a High level.
OFF/LOW This channel is active when either the switch input is Off (or opened), or the logic level voltage
input is at a Low level.
4. Select a warning option and push enter.
Option
Description
OFF
An active discrete input does not trigger a device warning.
ON
An active discrete input triggers a device warning.
English 37
- AMTAX sc (118 pages)
- PHOSPHAX sc (106 pages)
- CL17 Instruction sheet (8 pages)
- CL17 USER MANUAL (60 pages)
- CLF10 sc (382 pages)
- 9184 sc (50 pages)
- POLYMETRON 9611 sc Installation (398 pages)
- POLYMETRON 9610 sc Maintenance and Troubleshooting (286 pages)
- POLYMETRON 9611 sc USER INSTRUCTIONS: SEQUENCER LINE INSTALLATION (2 pages)
- POLYMETRON 9611 sc USER INSTRUCTIONS: SS SAMPLE CONDITIONING (4 pages)
- HACH 5500 sc Installation (408 pages)
- HACH 5500 sc Maintenance and Troubleshooting (286 pages)
- HACH 5500 sc PO43-HR Operations (390 pages)
- HACH 5500 sc PO43-LR Operations (392 pages)
- HACH 5500 sc SiO2 Installation (418 pages)
- HACH 5500 sc SiO2 Operations (394 pages)
- POLYMETRON 9610 sc SiO2 Installation (410 pages)
- POLYMETRON 9240 Basic User Manual (162 pages)
- POLYMETRON 9240 Operator Manual (144 pages)
- PHOSPHAX sigma Basic User Manual (305 pages)
- AMTAX inter2 Basic User Manual (377 pages)
- GANIMEDE P (60 pages)
- BODTrak II Basic User Manual (172 pages)
- BODTRACK II User Manual (28 pages)
- BODTRAK II: Respirometric Biochemical Oxygen Demand (BOD) (46 pages)
- SPECTRO COLOR d-8 (29 pages)
- REFO 60_REFO 60D (36 pages)
- REFO 3_REFO 3D (19 pages)
- LICO 620 Basic User Manual (329 pages)
- LICO 620 User Manual (114 pages)
- SD900 (2 pages)
- LZX971 (8 pages)
- SC Sensor Plug and SC Sensor Coupling (8 pages)
- HQ Series Portable Meters (4 pages)
- LDO Sensor Replacement Kit for use with Hach HQ10 and HQ20 (43 pages)
- LDO Sensor Replacement Kit (8 pages)
- SIP 10 (273 pages)
- SC 100 (58 pages)
- SI792 E_T Quick start guide (2 pages)
- SI792(x) E HACH GLI 3700 series SI792(x) T 7MA2200 and 8398 series User Manual (116 pages)
- SI792 D Quick start guide (2 pages)
- SI792 D, SI792x D, SI792x D-FF, SI792x D-PA User Manual (98 pages)
- SI794 C and SI794 D Short instructions (2 pages)
- SI794 D tr User Manual (108 pages)
- SI792 C Quick start guide (2 pages)