Modbus modbus register map – Hach-Lange SC 200 Modbus Module User Manual User Manual
Page 13
Modbus modbus register map
Group name
Tag name
Register # Data type
Length R/W Discrete
Range
Min/Max range Description
Setup
Baud rate
40001
Unsigned integer 1
R/W 0/1/2/3/4
Baud rate selection (0=9600; 1=19200;
2=38400; 3=57600; 4=115200)
Setup
Modbus mode
40002
Unsigned integer 1
R/W 0/1
Modbus mode (0=RTU; 1=ASCII)
Setup
Data order
40003
Unsigned integer 1
R/W 0/1
Register Data order (0=Little Endian
register order; 1=Big Endian register
order)
Setup
Parity
40004
Unsigned integer 1
R/W 2/0/1
Modbus parity (0=Even; 1=Odd;
2=None)
Setup
Stop bits
40005
Unsigned integer 1
R/W 1/2
Number of stop bits (1 or 2)
Setup/Addresses
Network card address
40006
Unsigned integer 1
R/W
0/246
Modbus address for the Modbus card
(1 to 246)
Setup
Modbus card name
40007
String
8
R/W
Location string of the network card
Network Timing
READ TIMEOUT
40015
Unsigned integer 1
R/W
1000/30000
Register read timeout setting (ms)
Network Timing
REG WRITE TMO
40016
Unsigned integer 1
R/W
3000/30000
Register write timeout setting (ms)
Network Timing
FILE WRITE TMO
40017
Unsigned integer 1
R/W
5000/30000
File write timeout setting (ms)
Network Timing
FILE PREP TMO
40018
Unsigned integer 1
6000/30000
File write timeout setting (ms)
Setup/Addresses
Device address
40019
Unsigned integer 1
R/W
0/246
The selected Modbus address of the
device (1 to 246)
Setup/Addresses
Select device
40020
Unsigned integer 1
R/W
0/30
Select the device to view/set Modbus
address (1 to 30)
Diagnostics
FUNCTION CODE
40021
Unsigned integer 1
R/W
0/65535
Function code used in the menu
system
Diagnostics
NEXT STATE
40022
Unsigned integer 1
R
0/65535
Next state value used in the menu
system
Diagnostics
INT BAUD RATE
40023
Unsigned integer 1
R
0/1/2/3/4
Baud rate selection (0=9600; 1=19200;
2=38400; 3=57600; 4=115200)
English 13