Troubleshooting
77
View sensor 
monitor
From the measuring mode press
CONF
and enter passcode
‘2222’ for the validation of the sensor and a complete 
measured-value processing. The measured resistance is 
shown in the main display and the measuring temperature is 
shown in the lower display. Press 
ENTER
to return
immediately to the measuring mode.
View the last error 
message
From the measuring mode, press
CONF
and enter passcode
‘0000’. The last error message will show for 20 seconds, or 
press 
ENTER
to return immediately to the measuring mode.
After viewing the error, the message will be deleted.
Specify output 
current
From the measuring mode, press
CONF
and enter passcode
‘5555’. The measured current will show in the secondary 
display. The output current in the main display can be 
modified. 
To change the output current: 
1. Select the current value with the 
RIGHT ARROW
key.
2. Edit the number using the
UP ARROW
key.
3. Press
ENTER
to confirm. The entered value will show in
the secondary display. The transmitter will remain in Hold 
mode. 
4. To exit Hold mode, press
CONF
, then
ENTER
to return to
measuring mode.
10.4 Diagnostic tests (continued)
Function
Description