beautypg.com

2 profibus pa communication, Section 7.2 on, Operation—profibus pa – Hach-Lange SI792 C, SI792x C, SI792x C-FF, SI792x C-PA User Manual User Manual

Page 65: 2 configuration menu (continued)

background image

Operation—Profibus PA

63

7.2 Profibus PA communication

Profibus uses a master/slave data exchange technique. The master
(typically a PLC) generates queries to individual slaves. The slaves, in
turn, reply back with a response to the master. A Profibus message
contains the information required to send a query or request, including
the slave address, function code, data, and a checksum. See

Table 13

on page 64

for Profibus communication parameters in case of errors.

Select temperature compensation

1

tc

OFF

Temperature compensation turned off

LIN (default)

Linear temperature compensation
Enter temperature coefficient:

Range: 0–19.99%/K (default: 2.00%/K)
Reference temperature = 25 °C

Enter temperature coefficient
02.00%/K (default) (xx.xx%/K)

nLF

Natural waters (to EN 27888)

–01– FCT

Ultrapure water with NaCl traces

–02– FCT

Ultrapure water with HCl traces

–03– FCT

Ultrapure water with NH

3

traces

Select Sensocheck

CHECK ON

Continuous Sensocheck evaluation of
sensor function

CHECK OFF (default)

Enter Profibus address

2

Edit

0126 BUS (default)

Range: 0001–0126

1

Temperature compensation is not available for salinity.

2

Use only when there is no bus connection. The transmitter will restart and set all

parameters to default values. Individual settings must be entered once more.

7.1.2