Specifications – Hach-Lange POLYMETRON 9240 Operator Manual User Manual
Page 16

14
Specifications
MECHANICAL SPECIFICATIONS
Maximum panel dimension
(H x L x D)
850 x 450 x 252.5mm [33.46 x 17.71 x 9.94in]
Inlet
Simple fittings for 6 mm O.D. tubing or ¼" O.D. in
PE-low density. ¼" OD in PHED-PTFE-SS as option
Outlet
Barbed stem for 12 mm (½" I.D.) hose
Protection rate
Transmitter: IP65 (NEMA 4)
Panel: IP50 (Dust protection)
Optional Enclosure: IP54 (Splash water proof),
Instrument is designed to avoid DIPA vapor inside the enclosure.
All DIPA vapor is collected and sent to the instrument drain
Cell
PMMA - compact (minimum tubing)
Flame rate
Conform UL
Maximum weight
15 - 30 Kg
INTERFACE SPECIFICATIONS
Mains power supply
100 - 240 VAC, 50-60 Hz, ± 10%, automatic switching
Max. consumption
80 VA.
Fuse
5 x 20 cartridge T2AL-250V following CEI127
Display
Curve trend, Last Cal Date, Historical, Concentration,
Temperature, Potential
Analog outputs
Number: 6
4-20 or 0-20 mA (650 ohms)
Linear / Dual / Logarithm
Event indication
Relays
4 x Relay (conc)
1 x Warning
1 x System
Logic input
Active / Inactive channels
Remote AutoCal
STANDARDS
European standards
EN 61326 Class A for EMC
EN 601010-1 for low voltage safety
International standards
cETLus
1