Section 2 specifications, 1 technical specifications – Hach-Lange POLYMETRON 9240 Operator Manual User Manual
Page 15
13
Section 2
Specifications
2.1 Technical specifications
Specifications are subject to change without notice.
Table 1 Technical specifications
PERFORMANCE SPECIFICATIONS
Measuring range
0 to 10,000 ppb freely programmable
0 to 200 ppm with K-Kit option
Accuracy
Non-cationic application:
± 0.1 ppb or ± 5% of reading, whichever is greater
Cationic application:
± 2 ppb or 5% reading, whichever is greater
Repeatability
< 0.02 ppb or 1.5% reading, whichever is greater within 10°C
variation
Average response times at 25°C
with a maximum
ΔT =15°C between
channels
T90%
≤ 10 min
Concentration step
from one channel to
another
Max. temp
variation
(°C)
Time to reach accuracy 0.1
ppb or 5%
up
down
0.1
↔ 5 ppb
3
9 min
27 min
0.1
↔ 50 ppb
3
11 min
41 min
0.1
↔ 200 ppb
3
9 min
45 min
< 0.1
↔ 1 ppb
(1)
3
29 min
36 min
0.1
↔ 50 ppb
15
11 min
41 min
Electrode type
pH glass electrode
Number of channels
1 to 4
Interference phosphate 10 ppm
Measurement variation less 0.1 ppb.
Sample temperature interference
< 0.5% / °C
ENVIRONMENTAL REQUIREMENTS
Typical environment
Power station / indoor / demineralized water plant or
instrumentation room
Suspended solids
< 2 NTU, no oil, no grease.
Temperature range for storage
-20 to 60°C (2 to 140°F)
Relative humidity
10 to 80%
Ambient temperature
5 to 50°C (41 to 122°F)
Sample temperature variation
Stabilization in 10 mins from 15°C to 30°C
Use the static heat exchanger system when the temperature
difference between samples is higher than 15°C
pH range of sample
Non-cationic application: 6 to 10 pH
Cationic application: 2 to 10 pH
Flow rate
6 to 9 L/hour
Pressure
0.2 to 6 bar (3 - 87 psi)
Acidity
Less than 250 ppm (equivalent CaCO
3
)
Power supply voltage fluctuation
± 10%
Over voltage category
2 (according to standard EN 61010-1)
Pollution degree
2 (according to standard CEI 664)
Altitude
< 2000 m
Measurement category
Cat II, Class 1 (overvoltage < 1500V)