Hach-Lange HACH 5500 sc SiO2 Installation User Manual
Page 14
W A R N I N G
Electrical shock and fire hazards. Make sure that the supplied cord and non-
locking plug meet the applicable country code requirements.
W A R N I N G
Electrocution hazard. Make sure that the protective earth conductor
has a low impedance connection of less than 0.1 ohm. The connected
wire conductor must have the same current rating as the AC mains line
conductor.
N O T I C E
The instrument is used for a single phase connection only.
Cord installation: The manufacturer recommends to use the optional
cord and sealing gland. Refer to the maintenance manual for the
replacement parts list. For a customer-supplied cord, three 18-gauge
conductors are required and the cord must be shorter than 3 meters
(10 feet). Use a sealing type strain relief to keep the environmental rating
of the instrument. Refer to
on page 3. To connect power
or
and
.
Table 3 AC wiring information (AC models only)
Terminal
Description
Color—North
America
Color—EU
1
Protective Earth (PE)
Ground
Green
Green with
yellow stripe
2
Neutral (N)
White
Blue
3
Hot (L1)
Black
Brown
Table 4 DC wiring information (DC models only)
Terminal
Description
Color—North
America
Color—EU
1
Protective Earth (PE)
Ground
Green
Green with
yellow stripe
2
24 VDC return (–)
Black
Black
3
24 VDC (+)
Red
Red
14 English