beautypg.com

1 probe assembly, Figure 5, 4 connecting the waste stream – Hach-Lange 9187 sc User Manual

Page 13: 5 assembling and placing the probe, Installation

background image

11

Installation

3.4 Connecting the Waste Stream

Connect the waste stream using the supplied ½-inch ID tubing. Be sure the drain is free
flowing (free of obstructions) so that the waste stream does not cause unnecessary
back-pressure or overflow.

Note: Waste from this instrument must go to the drain.

3.5 Assembling and Placing the Probe

Refer to

Figure 5

for a detailed description of the probe components.

Figure 5

Probe Components*

3.5.1 Probe Assembly

CAUTION

To familiarize yourself with handling precautions, dangers and emergency
procedures, always review the Material Safety Data Sheets prior to handling
containers, reservoirs, and delivery systems that contain chemical reagents and
standards. Protective eye wear is always recommended when contact with
chemicals is possible.

1. Screw the membrane cap onto the probe body (

Figure 6

and

Figure 7

). Be careful not

to touch or damage the membrane surface.

2. Remove the filling screw from the probe body.

3. Visually inspect the electrolyte to ensure that there are no particles or other

impurities present.

4. Using the supplied syringe, fill the probe body with ~7 mL of electrolyte.

5. Slowly insert the electrode into the probe body. Do not use force when inserting into

the probe body.

1.

Electrode Retaining Ring

2.

Measurement Electrode

3.

Probe Body

4.

Filling Screw

Note: A small hole exists in the plug to allow the instrument to maintain a consistent pressure regardless of changes in
atmopsheric pressure.

5.

Flange

6.

Probe Body Washer

7.

Pre-Mounted Membrane (set of four); Ensure the properly marked membrane is chosen (e.g. CL will appear on the side
for a Chlorine membrane).

*See

Replacement Parts and Accessories on page 31

.

This manual is related to the following products: