Flowline LU84-5161 EchoSpan User Manual
Page 21

31 MAR 11
EchoSpan
21 of 22
Rev B
MN301550
Appendix:
Setup:
You can view how the EchoSpan is configured.
1. From the main MENU level, press SETUP when VALUES appears.
2. When SETUP appears, press the SELECT key.
3. Setup will display the following information:
a. Units, Display, Rev mA, Safe, Height, Fill-H
4. When completed, press SELECT when EXIT appears to return to the main program level.
Diagnostics (DIAG) Parameters:
This mode runs diagnostic tests that confirm operation of EchoSpan. This mode should only be
used when supervised by a Flowline representative.
Reset:
EchoSpan enables the end user to reset the entire configuration back to the original factory settings.
Follow the instructions below to reset EchoSpan:
1. From the main MENU level, press SELECT when VALUES appears.
2. When RESET appears, press the SELECT key.
3. When YES appears, press SELECT key to reset EchoSpan.
a. To cancel the reset, press SELECT when NO appears.
4. When completed, press SELECT when EXIT appears to return to the main program level.
Factory Settings:
EchoSpan
HEIGHT
FILL-H
LU80 Series
118.1” (300 cm)
114.1” (290 cm)
LU81 Series
216.5” (550 cm)
208.5” (530 cm)
LU83 Series
314.9” (800 cm)
307.1” (780 cm)
LU84 Series
393.7” (1000 cm)
381.9” (970 cm)