Operating instructions – Fill-Rite 900 Series Digital Liquid Meter User Manual

Page 5

background image

5

1) Pressing and holding the CAL

button for 2 seconds before
dispensing into a container will
cause the meter to skip a fill
operation and only change the
unit of measure.







1) The left-most digit will begin

blinking.

2) Use the TOTAL button to scroll from 0 to 9 until the desired number has been

selected.

3) Press and release the CAL

button to move to the next digit.

4) Repeat Steps 2 - 4 until the

amount of dispensed fluid has
been entered.

5) If you make a mistake entering a

digit, press and release the CAL
button repeatedly until the digit
is active again.

6) Press and hold the CAL button

for 2 seconds to complete the
calibration process regardless of
the current digit selected.

7) If at the last digit and the CAL

button is pressed for less than 2 seconds, the process should roll back to the left-most
digit, but retain the current value. This allows the user to edit an incorrect value.

8) An error check for gross calibration errors (+/- 15%) will be performed unless USER

units are selected. If an error is detected, the meter will display Err0 until a button is
pressed and the calibration process is aborted.

9) If the meter is left untouched for 2 minutes or the RESET button is pressed, the unit

will automatically exit Calibration Mode and ignore any changes.

Operating Instructions

For accurate measurement and to
prevent meter damage, dispenser and
piping must always be filled with liquid
and be free of air. The meter portion of
your dispenser should be calibrated per
instructions in this manual prior to its
use.

The Fill-Rite 900 Series Digital Meter
face incorporates three buttons that
control all the operation and calibration
functions.

To turn the display on, or wake the
meter from “Sleep” mode, simply press
any button on the face. The display will
turn on and you can proceed with the
function you desire. Dispensing fluid will also turn the meter display on.

IMPORTANT! For maximum
accuracy, try to get as close as
possible to the dispensed
volume without changing flow
rate
.