Rev. 12/97 -30- D77X

useful for accommodating a very large accumulated flow.
The exponent is a “ X 10

n

” multiplier, were “n” can be from

0 (10

0

, X 1 multiplier) to 4 (10

4

, X 10,000 multiplier).

The External Counter output, available at the two
terminals labeled CTR on the Main PCB, is influenced by
the TOTAL MULT value. Since the output is designed to
operate electromechanical accumulators, large flow rates
will require that the TOTAL MULT be set to a value usable
by these types of counters (typically speeds no faster than
3 cps). The following chart tabulates suggested settings

vs. flow ranges:

TOTAL ON/OFF

Exponent

Multiplier

Useable CTR
Range:

GPM or LPM

0

X1

30-600

1

X10

300-6,000

2

X100

3,000-60,000

3

X1,000

30,000-600,000

4

X10,000

300,000-6,000,000

Key press number

Operation

First press

Stops the internal totalizer/
external CTR and displays
the last value

Second press

Resets the internal totalizer
to zero

(continued)

PART 3 - KEYPAD CONFIGURATION