Part 2 - transducer installation – Dynasonics 902_903 Series Portable Doppler Ultrasonic Flow Meter User Manual

Page 15

background image


Rev. 10/08

- 13 -

D902/3

PART 2 - TRANSDUCER INSTALLATION



Before the transducer heads are bonded to the pipe
surface, an area slightly larger than the flat surface of the
transducer must be cleaned to bare metal on the pipe.
(Plastic pipes do not require preparation beyond removal
of paint.) Remove all scale, rust and paint. Thoroughly
dry and degrease the mounting surfaces.

NOTE: Small pits in the piping surface typically do not
significantly impact ultrasonic transmission or signal
reception.


After selecting the applicable mounting location and
preparing the piping surface as detailed in Steps A and B,
the transducer can be mounted to the pipe.


Step B -
Pipe Surface
Preparation

Step C -
Transducer
Mounting

NOTE: High Temperature transducer installations require
specialized mounting hardware and instructions. For
drawings detailing installation of this option, contact
the factory.

Steps A and B, Mounting Locations and Pipe Preparation
sections of this manual apply to the High Temperature
option. Reference these sections as required.

Figure 6