6 Maintenance
WARNING
The following instructions are for use by qualified personnel only. To avoid electrical
shock, do not perform any servicing other than contained in the operating
instructions unless you are qualified to do so.
6.1 Fuse Replacement
If the fuse blows, the CV or CC indicators will not light and the power supply will not
operate. The fuse should not normally open unless a problem has developed in the
unit. Try to determine and correct the cause of the blown fuse, then replace only
with a fuse of the correct rating as listed below. The fuse is located on the rear panel
(see
Line Voltage
Fuse
Type
115 Vac ± 10%, 50/60 Hz
6A
Slow Blow
230 Vac ± 10%, 50/60 Hz
3A
Slow Blow
Table 2 - Fuse Values
Line Voltage Conversion, International Units
The primary winding of the power transformer is tapped to permit operation from
120/220/230 or 240 VAC, 50/60 Hz line voltage. Conversion from one line voltage to
another is done by a simple position change of the fuse receptacle underneath the
line cord.