Multi 1970i

Operation

39

4.5.4

AutoRead AR (Drift control) and hold function

The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values. With the aid of the hold function the
measured value display is frozen.

For conductivity measurements, use the AutoRead function and hold
function like this:

Note
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing . You can
only change to another measuring mode after completion of AutoRead.

1

Call up the measuring mode

or SAL with and/or <

>

<

>.

2

Immerse the conductivity measuring cell in the test sample.

3

Activate the AutoRead function with . The cur-
rent measured value is frozen (hold function).

4

Start AutoRead with .
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.

5

If necessary, start the next AutoRead measurement with
.

6

To terminate AutoRead: Press the key.