Commissioning

Multi 1970i

14

3.4

Sensor quiver

To store the sensors during field operation and to keep the sensor ele-
ment moist, the quiver tip contains a sponge rubber insert that can be
moistened with deionized water.

Note
For further details on proper storage, refer to the operating manual of
the sensor.

Moistening the

quiver insert

Warning
Do not store pH electrodes in the quiver for more than 10 hours. To
store them for a longer period of time, use the watering cap filled with
potassium chloride (3 mol/l) of the electrode.

1

Press the quiver out of the holder from the back side of the in-
strument and pull it out completely.

2

Pull off the quiver tip and moisten the sponge rubber with
deionized water.