Multi 1970i

Operation

37

Temperature

compensation

The nonlinear temperature compensation is set fixed and is shown on
the display by nLF.

Reference temperature,

Tref

The reference temperature (Tref) can be switched between 20 °C and
25 °C. It appears on the display as Tref20 or Tref25. To switch over the
reference temperature, see S

WITCHING

OVER

THE

REFERENCE

TEMPERA

-

TURE

, page 60.

Checking the

cell constant

1

Press the key repeatedly until CAL disp appears on the
display.

2

Press the repeatedly until the last calibrated
cell constant is displayed, e. g. 0.472 1/cm.

3

To return to the measuring mode: Press the key when the
correct cell constant is displayed.

4

If you want to recalibrate the cell constant, proceed according
to section 4.5.5 D

ETERMINING

THE

CELL

CONSTANT

(C

ALIBRATION

IN

THE

CONTROL

STANDARD

) .