Xylem WQ pH SENSORS User Manual

Page 7

background image

Global Water

800-876-1172

globalw.com

- 7 -

V.

pH Sensor

a. pH sensor specifications.

Output:

4-19mA

Range: 0-14

pH

Accuracy:

2% of full scale

Operating Voltage:

10-36VDC

Current Draw:

16.6 mA plus sensor output

Warm Up Time:

3

seconds

minimum

Operating Temperature: -5° to +55°C
Pressure Rating:

0-100 psi

Size of Probe:

1 1/4" diameter x 10" long

Weight:

1 lb.

b. The sensing element is covered with a protective cap while at the

Global Water facility to prevent the sensor from becoming damaged.
This cap must be removed prior to sensor installation or the readings
will be in error. To remove the cap, remove the protective shield and
remove the rubber cap covering the sensing element. Replace the

protective shield. Note: Save the cap for future use.


c. This sensor has a removable sensing electrode. If the sensor is

reading incorrectly, after following the basic maintenance steps, the

electrode should be removed and the metal contacts cleaned. If this
does not improve the sensor’s results the electrode should be
replaced. To remove the sensing electrode loosen the set screw
holding the sensor shield and remove the shield from the sensor.

Unscrew the sensing electrode from the sensor housing. Clean the
metal contacts of the electrode with a clean cloth. If the electrode is
replaced with a new electrode the sensor must be recalibrated to work

correctly.


d. The pH sensor is a three-wire sensor. Three wire sensors use the red

wire for positive voltage, the white wire for the output signal, and the

black wire for ground. Warning: Always connect the sensor with
the power turned off.