pH Calibration check
Step 1) Fill one container with tap water and another with 4 pH buffer

solution.

Step 2) Place the pH sensor into the container with 4 pH buffer. Turn

on the power supply and the current meter. Let the sensor

stabilize for 5 minutes before taking any measurements.

Step 3) Record the output current of the sensor, W = _______.

Remove the sensor and rinse it off in the tap water.

Step 4) Fill a container with 10 pH buffer solution.

Step 5) Place the pH sensor into the container. Turn on the power

supply and the current meter. Let the sensor stabilize for 5
minutes before taking any measurements.

Step 6) Record the output current of the sensor, X = _______. Remove

the sensor and rinse it off in the tap water.

Step 7) Subtract 4 pH current output from the 10 pH current output

,

X -

W

= _______ = C.

Step 8) Calculate B. X – (C/6)(10)= _______ = B.

Step 9) Find the low current value for the sensor. B= _____ = L

C

. This

current is the output current the sensor would produce if the pH
were 0.

Step 10) Find the high current value for the sensor. (C/6)(14)+ B=

_____ = H

C

. This current is the output current the sensor would

produce if the pH were 14.

Step 11) Use these new current values to recalibrate the system that is

monitoring the sensor output.