Mounting on a pipe stub, Mounting – Xylem PHIX COMPACT pH and Redox Transmitter User Manual

Page 12

background image

12

Mounting

GB pHix Compact Manual 100318

Mounting on a pipe stub
In order to ensure a tight seal and correct
function of pHix

®

Compact, the following

should be complied with:
1: The pipe stub must be flush and free of

scales, burrs etc.

2: The pipe stub must have a length so

that the tip of the electrode is exposed
approximately 10 mm into the liquid.

3: Take care to have adequate room for

mounting and removal o f pHix

®

Com-

pact.

4: The working pressure of the system

must not exceed 5 bar g.

5: The working temperature of the s ystem

must not exceed 80 °C.

See also the following illustrations for correct
adaptation to the pipe system.

Approx. 10 mm

pHix

®

Compact is mounted correctly when the

electrode tip is exposed approximately 10 mm out

from the inner lining of the pipe.

Min. 25 mm
thread

2 " outer pipe
thread (ISO
228-1)

80 mm to

inner lining

Viton gasket

Min. 320 mm
free space for
mounting/re-
moval of
pHix

®

Compact.

Max. working pressure = 5 bar g.

Maks. working temperature = 80 °C.