Mounting on a pipe stub, Mounting – Xylem PHIX COMPACT pH and Redox Transmitter User Manual
Page 12

12
Mounting
GB pHix Compact Manual 100318
Mounting on a pipe stub
In order to ensure a tight seal and correct
function of pHix
®
Compact, the following
should be complied with:
1: The pipe stub must be flush and free of
scales, burrs etc.
2: The pipe stub must have a length so
that the tip of the electrode is exposed
approximately 10 mm into the liquid.
3: Take care to have adequate room for
mounting and removal o f pHix
®
Com-
pact.
4: The working pressure of the system
must not exceed 5 bar g.
5: The working temperature of the s ystem
must not exceed 80 °C.
See also the following illustrations for correct
adaptation to the pipe system.
Approx. 10 mm
pHix
®
Compact is mounted correctly when the
electrode tip is exposed approximately 10 mm out
from the inner lining of the pipe.
Min. 25 mm
thread
2 " outer pipe
thread (ISO
228-1)
80 mm to
inner lining
Viton gasket
Min. 320 mm
free space for
mounting/re-
moval of pHix
®
Compact.
Max. working pressure = 5 bar g.
Maks. working temperature = 80 °C.
- H-2191 (57 pages)
- H-223 (9 pages)
- H-264 (16 pages)
- H-310 (40 pages)
- H-312 (38 pages)
- H-3301 (24 pages)
- H-3342 (30 pages)
- H-335 (32 pages)
- H-339 (21 pages)
- H-340-CalKit (6 pages)
- H-340 (12 pages)
- H-340SDI (32 pages)
- H-3521 (57 pages)
- H-3531 FlashLite (76 pages)
- H-3551 (42 pages)
- H-3553 (36 pages)
- H-375 04101 WIND MONITOR-JR (18 pages)
- H-4161 (36 pages)
- H-4191 (21 pages)
- H-424MS V1.0 (25 pages)
- H-424MS V1.1 (22 pages)
- H-4271 (22 pages)
- H-4280 (34 pages)
- H-4400 (7 pages)
- H-4401 (6 pages)
- H-5223 (107 pages)
- STORM 3 UPGRADE (4 pages)
- System 5000 BASIC Manual (54 pages)
- System 5000 User Manual (136 pages)
- DH-21 (112 pages)
- BrazedPlate Heat Exchangers (4 pages)
- Heat Exchangers (8 pages)
- GL500 MULTICHANNEL DATA LOGGER (32 pages)
- 4046 ANALOG TO SDI-12 CONVERTER (7 pages)
- 4015 SDI-12 TO ANALOG CONVERTER (5 pages)
- iRIS-CAM Wastewater Security Camera (11 pages)
- SDI-12 Data Logger iRIS 220, iRIS 320, iRIS 350, iRIS 150, iRIS 300 (111 pages)
- iRIS 150 GPRS DATALOGGERS (53 pages)
- iCE3 GPRS Variant (29 pages)
- GL500-2-1 Data Logger (32 pages)
- 3101 CONDUCTIVITY CONTROLLER (11 pages)
- 3675 pH CONTROLLER (9 pages)
- 378 TEMPERATURE CONTROLLER (9 pages)
- Expert 1400 (16 pages)
- Expert 700 (8 pages)