3.1 Configuring the RS–485 Port (continued)
For Modbus some applications, especially involving the ASCII operation mode, it may be
necessary to set a different protocol than the default setting (8 bits, 1 stop bit, no parity).
Menus are available for this in the Extended Settings portion of the configuration
CONFIG mode.

While in the CONFIG mode, press the

↵ button, several times until the Extended Settings

is as shown below. Select On using the

t or u buttons.










Press the

button a few times until the menus appear for BITS, PRTY (Parity) and

STOP (Stop Bits). Set each one to the desired setting using the

t andu buttons.