Address Type

Register

Min

Default

Max

Function
2 -- Odd

40013

Int

RS-485 stop bits

0

0

1

0 – One
1 -- Two

40014 Int

Instrument
address

1 1

255

40015 Int

Modbus

serial

mode

0

0

1

0 – RTU
1 – ASCII

40016

Int

Alarm 1 type

0

0

2

0 – Off
1 – Low alarm
2 – High alarm
3 – Error alarm

40017, 40018

Float Alarm 1 set point

0.0

1.0

10.0

40019

Int

Alarm 2 type

0

0

2

0 – Off
1 – Low alarm
2 – High alarm
3 – Error alarm

40020, 40021

Float Alarm 2 set point

0.0

1.0

10.0

40022, 40023

Unused

40024 Int

Measurement
period

90 150 600

Seconds

between

measurements

40025 Int

Water
Conservation

0 0 1

Water

Conservation

Flag

40026

Unused

CL500 Modbus (3/09)
Rev. 1.3