Appendices – Xylem DO600 Dissolved Oxygen Meter User Manual
Page 9

DO600 - Version 5.0 8/05
9
Appendices
Troubleshooting Guide
Symptom Possible
Causes
Action
Unit will not power on
-Batteries not in place
-Dead batteries
-Battery polarity incorrect
-Replace Batteries
-Replace Batteries
-Reorient/Replace Batteries
“BAT” indicator shown on display
-Batteries are weak
-Replace Batteries
Unstable Readings
-Insufficient electrolyte in probe (Air
bubbles present when probe is
inverted.)
-Electrolyte is depleted
-Replace electrolyte, and
membrane cap assembly.
Readings drift down
-Insufficient stirring (the probe
consumes oxygen at the measuring
surface, requiring constant sample
movement across membrane)
-Move probe in sample or stir
sample
Slow Response
-Dirty or damaged membrane
-Replace electrolyte and
membrane cap
Electrode cannot be calibrated
-Depleted electrolyte
-Dirty or damaged membrane
-Replace electrolyte and
membrane cap
Electrode cannot be calibrated
after replacing the electrolyte and
membrane cap
-Dirty probe (Cathode is not a shiny
gold color)
-Clean cathode with cleaning
paper
Sample reading is frozen
-Unit is in “HOLD” mode
-Unit is locked
-Release HOLD (momentarilly
press Mode/Hold button)
-Remove batteries, press
ON/OFF button, replace batteries
and restart