Troubleshooting Guide

Symptom Possible

Causes

Action

Unit will not power on

-Batteries not in place
-Dead batteries
-Battery polarity incorrect

-Replace Batteries
-Replace Batteries
-Reorient/Replace Batteries

“BAT” indicator shown on display

-Batteries are weak

-Replace Batteries

Unstable Readings

-Insufficient electrolyte in probe (Air
bubbles present when probe is
inverted.)
-Electrolyte is depleted

-Replace electrolyte, and
membrane cap assembly.

Readings drift down

-Insufficient stirring (the probe
consumes oxygen at the measuring
surface, requiring constant sample
movement across membrane)

-Move probe in sample or stir
sample

Slow Response

-Dirty or damaged membrane

-Replace electrolyte and
membrane cap

Electrode cannot be calibrated

-Depleted electrolyte
-Dirty or damaged membrane

-Replace electrolyte and
membrane cap

Electrode cannot be calibrated
after replacing the electrolyte and
membrane cap

-Dirty probe (Cathode is not a shiny
gold color)

-Clean cathode with cleaning
paper

Sample reading is frozen

-Unit is in “HOLD” mode
-Unit is locked

-Release HOLD (momentarilly
press Mode/Hold button)
-Remove batteries, press
ON/OFF button, replace batteries
and restart