Maintenance, cleaning, replacement

CellOx 325

24

5.3

Changing the electrolyte and membrane head

Caution

Before starting to work with the sensor, please note the G

ENERAL

MAINTENANCE

INSTRUCTIONS

on page 23.

General

information

WTW delivers the sensor ready to operate (see section 3). The electrolyte
solution and the membrane head must only be replaced if:

l

a calibration error occurs and the membrane is heavily contaminated

l

the membrane is damaged

l

the electrolyte solution is exhausted

l

in case of a leak message by the measuring instrument.

Changing the electrolyte solution and the membrane head

Unscrew the membrane
head.

Caution: Electrolyte solution!
For disposal of the membrane
head and electrolyte solution,
see section 5.6.

Rinse the sensor head with
deionized water.

Carefully rub and dry the
counter electrode with a lint-
free paper towel.

Immerse the sensor head in-
cluding the counter electrode
in RL/G cleaning solution.

Allow to react for 1 to 3 min-
utes.

Thoroughly rinse the sensor
head with deionized water.

Water the counter electrode in
deionized water for at least
10 minutes.

RL/G

H O

2