CellOx 325

Maintenance, cleaning, replacement

23

5

Maintenance, cleaning, replacement

5.1

General maintenance instructions

For your safety

When dealing with the ELY/G electrolyte solution, observe the following
safety instructions:

Caution

The ELY/G electrolyte solution irritates eyes and skin. When dealing
with the ELY/G electrolyte solution, observe the following points:

l

During working activities, always wear suitable protective gloves
and protective goggles/face shield.

l

If it comes into contact with the skin, rinse thoroughly with water and
immediately change contaminated clothing.

l

If it comes into contact with the eyes, rinse thoroughly with water
and consult a doctor.

l

Follow the safety datasheet.

Caution

Before all maintenance activities, disconnect the sensor from the
instrument.

Note
Information on how to order wear parts and maintenance accessories can be
found in chapter 8 W

EAR

PARTS

AND

ACCESSORIES

.

5.2

Exterior cleaning

Cleaning agents

After cleaning, thoroughly rinse with deionized water and recalibrate if
necessary.

Contamination

Cleaning procedures

Lime sediments

Immerse in acetic acid for 1 minute (volume
share = 20 %)

Fat/oil

Clean with warm water that contains wash-
ing-up liquid