One channel temperature probe

Probe TPC 300 (NTC) for EBI 300

Measurement range

-35

°C…+75°C

Accuracy

+/-0.5°C

( -2

0°C…+40°C)

+/- 0.8°C (else)

One channel temperature probe

Probe TPX 220 (PT 1000) for EBI 310

Measurement range

-200

°C…+250°C

Accuracy

+/-2.0°C

( -20

0°C…-100°C)

+/- 1.0°C

( -10

0°C…-20°C)

+/- 0.2°C

( -20

°C…+60°C)

+/- 0.5°C

( +60

°C…+250°C)

One channel temperature probe

Probe TPX 250 (PT 1000) for EBI 310

Measurement range

-85

°C…+50°C

Accuracy

+/-1.0°C

( -85

°C…-20°C)

+/- 0.2°C

( -2

0°C…+50°C)

Two channel temperature probe
TPX 310 (PT 1000) for EBI 310

Measurement range

-

200°C…+400°C

Accuracy

+/-2,0 °C

( -

200°C…-100°C)

+/- 1,0°C

( -

100°C…-20°C)

+/- 0,2°C

( -

20°C…+60°C)

+/- 0,5°C

( +60°C…+250°C)

+/- 1,0°C

(+250°C…+400°C)

Capteur de température avec une voie
Sonde TPC 300 (NTC) pour l´ EBI 300

Plage de mesure

-35

°C…+75°C

Précision

+/-0,5°C

( -2

0°C…+40°C)

+/- 0,8°C (sinon)

Capteur de température avec une voie

Sonde TPX 220 (PT 1000) pour l´ EBI 310

Plage de mesure

-200

°C…+250°C

Précision

+/-2,0°C

( -20

0°C…-100°C)

+/- 1,0°C

( -10

0°C…-20°C)

+/- 0,2°C

( -20

°C…+60°C)

+/- 0,5°C

( +60

°C…+250°C)

Capteur de température avec une voie

Sonde TPX 250 (PT 1000) pour l´ EBI 310

Plage de mesure

-85

°C…+50°C

Précision

+/-1,0°C

( -85

°C…-20°C)

+/- 0,2°C

(-2

0°C…+50°C)

Capteur de température avec deux voies
TPX 310 (PT 1000) pour l´ EBI 310

Plage de mesure

-

200°C…+400°C

Précision

+/-2,0 °C

( -

200°C…-100°C)

+/- 1,0°C

( -

100°C…-20°C)

+/- 0,2°C

( -

20°C…+60°C)

+/- 0,5°C

( +60°C…+250°C)

+/- 1,0°C

(+250°C…+400°C)

This manual is related to the following products: