beautypg.com

H-424ms – Xylem H-424MS V1.1 User Manual

Page 17

background image

2.5 Send Acknowledge Command

The Send Acknowledge Command returns a simple status response which includes the address of
the sensor. Any measurement data in the sensor's buffer is not disturbed.

Command

Response

-------------

-------------------------

"a!"

"a"

Where:

a

Is the sensor address ("0-9", "A-Z", "a-z", "*", "?").

2.6 Send Identification Command

The Send Identification command responds with sensor vendor, model, and version data. Any
measurement data in the sensor's buffer is not disturbed.

Command

Response

-------------

------------------------------------------------------------

"aI!"

"allccccccccmmmmmmvvvxx...xx"

Where:

a

is the sensor address ("0-9", "A-Z", "a-z", "*", "?").

I

is an upper-case ASCII character.

ll

is the SDI-12 version compatibility level, e.g. version 1.2 is represented as
"12".

cccccccc

is an 8 character vendor identification to be specified by the vendor and
usually in the form of a company name or its abbreviation.

mmmmmm

is a 6 character field specifying the sensor model number.

vvv

is a 3 character field specifying the sensor version number.

xx...xx

is an optional field of up to a maximum of 13 characters to be used for
serial number or other specific sensor information not relevant to operation
of the data recorder.

H-424MS