H-340SDI

SDI-12 Command and Response Protocol 2-11

Example of a H-340SDI Extended "Read Offset" command:

Command

Response

Time Values

Description

-------------------

------------------------ ------ ---------

-----------------------------

"XRO!"

"a0011"

1 sec 1

Read Offset

Subsequent Command

Response

---------------------------

------------------------------------

"aD0"

a+0.000

2.13 Extended “Write Offset” Command

The H-340SDI maintains an offset value in RAM which is automatically added to the Total

Accumulation value. The offset value can be used to match the H340's Total Accumulation to the

current water year rainfall. This command allows the user to write (change) the current offset. The

offset is automatically set to zero at power-up or with the “aXRA” command. To change the Total

Accumulation to a new value you must first reset the accumulators with the “aXRA” command,

then set the offset with the “aXWO” command.

The new value is stored in RAM within the sensor and is not preserved if the power is lost. Once

the new offset value is written to the RAM, a copy is sent to the sensor data buffer for verification.

This data can be viewed by using a subsequent "D" command. To verify the Offset any other time,

use the "Read Offset" command.

Command

Response

Description

-------------

-------------------------

-----------------------------

"aXWOddd!"

"a0011"

Write Offset

Where:

a

is the sensor address ("0-9", "A-Z", "a-z", "*", "?").

XWO

are upper case characters.

ddd

is the new offset value. The input format is

very flexible. Some examples are shown below.

20.0

0.195

-500

5.93E-4