beautypg.com

H-340 sdi – Xylem H-340SDI User Manual

Page 13

background image

H-340SDI

H-340SDI Introduction 1-7

Figure 2

1.10 Adjusting the Tipping Bucket Mechanism

Warning: DO NOT ADJUST THE BUCKET STOPS.

Adjusting the stop posts causes a false reading of equal amounts of water in each bucket. This is

due to the variation of distance the bucket center must travel after passing the bottom of the

funnel. Do not adjust the bucket stops in an attempt to balance the weight of the buckets. Instead,

the H-340SDI provides an

adjustable screw parallel to the buckets by which the bucket weight

balance can be accomplished. This eliminates the need to adjust the stops and compensates for

any inequality of weight due to manufacture and assembly, and gives equal travel to each bucket

as the center passes the flow of water.

The counter weight underneath the bucket mechanism can be adjusted (up or down,) to set the

bucket volume (# tips/unit rainfall). See Figure 2. The up/down counter weight adjust

instructions in steps 1 and 2 may seem to be intuitively backwards however, realize the center of

gravity of the bucket and counterweight is above the pivot.

Counter Weight

Lock Nut #1

Bucket Stop

Do Not Adjust

2.062

Balance Screw

Set Screw