1-6 H-340SDI Introduction

H-340SDI

Other related equations of interest:

RATE

RAINFALL

N

in

hr

inches

hr

=

×

min

min

60

RAINFALL

RAINFALL

N

inches

tip

inches

tips

=

1.9 User Calibration

The “smart bucket” processor uses a fixed polynomial to compute each bucket volume as a

function of rainfall rate. The factory calibration should be sufficiently accurate for all rainfall

rates from 0 to 25 inches/hour. If the user requires an exact calibration at a rate other than 4.0

inches/hour, or if the tipping bucket mechanism is damaged or changed the bucket mechanism

may need to be recalibrated. The procedure for adjusting the counterweight is included below.

Please note that the “smart bucket” feature is calibrated somewhat differently than a normal

tipping bucket device. With a simple tippling bucket, a known amount of water is passed

through the gauge and the number of bucket tips is recorded. The counter weight is then adjusted

for the desired number of tips. With the H-340SDI “smart bucket” feature, the number of bucket

tips is not used in the calibration, instead, the SDI-12 measurement value is used. This is

different because the H-340SDI's microprocessor is using a non-linear polynomial expression to

compute the volume. With the aid of the polynomial you can get an accurate field calibration

without using the constant flow apparatus which is used at the factory. Just remember to use the

SDI-12 “aM” measurement values and not the number of bucket tips.