Appendix D H-310 Extended Commands D-7
Command Response
Description
"aXSF!" "a
Set sensor measurement speed to fast
Where:
a
Is the sensor address (0-9,*).
XSF
Are upper-case ASCII characters.
Example of a "Set Sensor Measurement Speed Fast" command:
Command
Response
"aXSF!"
"a
The set sensor measurement speed fast command causes the sensor to execute low resolution
measurements and make data available in 1 second.
NOTE:
The fast measure mode is not a low power mode. The sensor remains powered at
all times in this mode. However, in this mode data can be collected at a two
second interval.
D.7.3 Set Sensor Measurement Speed Slow
Command Response
Description
"aXSS!" "a
Set sensor measurement speed to slow
Where:
a
Is the sensor address (0-9,*).
XSS
Are upper-case ASCII characters.
Example of a "Set Sensor Measurement Speed Slow" command:
Command
Response
"aXSS!"
"a
The set sensor measurement speed slow command causes the sensor to execute high resolution
measurements and make data available in 8 seconds. This is the default setting.