D-4 Appendix D H-310 Extended Commands
H-310
D.4 Write "User Units Offset" Command
Command Response Description
"aXWOddd!" "atttn
Where:
a
Is the sensor address (0-9,*).
XWO
Are upper-case ASCII characters.
ddd
Is a the new user offset value. The input format is very flexible. Some examples are shown
below.
20.095
0.195
7.984E+10
167.824E5
005.9357E-7
500
ttt
Is a three digit integer (000-999) specifying the maximum time, in seconds, the sensor will
take to complete the command and have data available in its buffer.
n
Is a single digit integer (0-9) specifying the number of values that will be placed in the data
buffer.
The Write User Offset Command loads the "Users Units Offset" coefficient term. The new value
is stored in non-volatile EEPROM within the sensor. With the new offset value now written into
EEPROM, a copy of this value is placed in the sensor data buffer for verification. The
verification is carried out by using the "D" command. To verify the "User Units Offset" at any
other time use the "Read User Units Offset" command.
Example of a "Write User Units Offset" command:
Command Response Time Values Description
"aXWO12.34!" "a0021
coefficient