beautypg.com

H-310 – Xylem H-310 User Manual

Page 36

background image

D-4 Appendix D H-310 Extended Commands

H-310

D.4 Write "User Units Offset" Command

Command Response Description

"aXWOddd!" "atttn" Write user units offset coefficient

Where:

a

Is the sensor address (0-9,*).

XWO

Are upper-case ASCII characters.

ddd

Is a the new user offset value. The input format is very flexible. Some examples are shown
below.

20.095
0.195
7.984E+10
167.824E5
005.9357E-7
500

ttt

Is a three digit integer (000-999) specifying the maximum time, in seconds, the sensor will
take to complete the command and have data available in its buffer.

n

Is a single digit integer (0-9) specifying the number of values that will be placed in the data
buffer.

The Write User Offset Command loads the "Users Units Offset" coefficient term. The new value
is stored in non-volatile EEPROM within the sensor. With the new offset value now written into
EEPROM, a copy of this value is placed in the sensor data buffer for verification. The
verification is carried out by using the "D" command. To verify the "User Units Offset" at any
other time use the "Read User Units Offset" command.

Example of a "Write User Units Offset" command:

Command Response Time Values Description

"aXWO12.34!" "a0021" 2 sec 1 Write user units offset

coefficient