11 indexes, 1 explanation of the messages, 1 error messages – YSI IQ SensorNet NitraLyt Sensor User Manual
Page 59: 11 indexes -1, 1 explanation of the messages -1, 1 error messages -1
NitraLyt
®Plus
700 IQ
Indexes
11 - 1
ba76004e01
01/2012
11
Indexes
11.1
Explanation of the messages
This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ S
ENSOR
N
ET
system for the NitraLyt
®Plus
700 IQ sensor.
Note
Information on
the contents and structure of the log book and
the structure of the message code
is given in the L
OG
BOOK
chapter of the IQ S
ENSOR
N
ET
system
operating manual.
Note
The last three digits of the message code identify the source of the
message:
523 = NitraLyt+700 IQ (armature / component class, ADA adapters)
396 = NitraLyt+ (nitrate sensor)
11.1.1 Error messages
Message code
Message text
EA1396
Meas. range exceeded or undercut
* Check process
* Select other meas. range
EA2523
Sensor temperature too high!
* Check process and application
EA3523
Sensor temperature too low!
* Check process and application
EAN396
Chloride measurement: range exceeded or undercut
* Check process
EC7523
Cl electrode could not be calibrated, nitrate measurement blocked,
nitrate measurement possible with manual Cl compensation and Cl
electrode removed
* Check calib. conditions and standard
* View calibration history
* Service sensor immediately
(see operating manual)