YSI IQ SensorNet NitraLyt Sensor User Manual
Page 30
Matrix adjustment, check and calibration
NitraLyt
®Plus
700 IQ
4 - 6
ba76004e01
01/2012
The history of matrix adjustments contains the following information:
Note
The history of matrix adjustments of the chloride electrode is not output
with the calibration history of the sensor (NitraLyt+). If necessary, you
can create a log book message with the calibration history by means of
the menu item, Calib. history Cl in the setting table. How to do this is
described on page 3-10. The log book message is given in the log book
(message code IC5396).
Date
Date of the matrix adjustment
S
Slope [mV] of the electrode.
The slopes are marked by a star (*) because they were not changed
by the matrix adjustment or reset to the default condition with the
electrode zeroing.
DV
Drift voltage [mV]
0 is displayed after an electrode zeroing.
Ref1
Reference concentration [mg/l]
Ref2
This column is of no importance for the matrix adjustment
Cl-
Measured or entered chloride concentration [mg/l]
P
Procedure (1 = matrix adjustment)
T
Temperature [°C]
R
Evaluation of the matrix adjustment
+ : Matrix adjustment successful.
The sensor measures with the new adjustment data.
? : Matrix adjustment unsuccessful.
Invalid adjustment data was discarded. Measurement is
continued with the last valid values.