beautypg.com

3 basic cleaning, Basic cleaning -3 – YSI IQ S ENSOR N ET NitraVis 70x IQ TS User Manual

Page 39

background image

NitraVis 70x IQ TS

Maintenance and cleaning

5 - 3

ba75950e02

11/2012

5.2.3 Basic cleaning

Steps of the basic

cleaning

After completing the cleaning procedure, screw in the blind plug or

compressed air connection.

1.

Take the sensor out of the test sample and remove any solid

matter deposits and incrustation of fouling matter manually

with a brush or sponge.

2.

Wash the sensor down with warm tapwater (30 - 50 °C).

Note: Do not use a pressure washer or steam blaster.

3.

If the sensor might be contaminated with pathogenic germs it

should be disinfected with a surface disinfectant. This also

applies to the cable and mounting parts.

Note: Follow the instruction manual of the disinfectant (reac-

tion time!).

4.

Remove the blind plug or unscrew the compressed air connec-

tion from the sensor.

5.

Clean the sensor with soapy water. When doing so, clean the

area of the measuring gap and the compressed air channels

very thoroughly.

The compressed air channels can be cleaned with the aid of

interdental brushes and rinsed with soapy water from a wash

bottle.

6.

Soak the flocked cleaning cards with soapy water and use

them to clean the measuring gap as shown in the following fig-

ure. Cut suitable strips off the cleaning card to clean the bottom

of the measuring gap.

Then rinse the sensor with running warm tapwater and finally

with a lot of ultrapure water.

7.

Finally wipe the measurement window with a lens tissue.