YSI IQ S ENSOR N ET NitraVis 70x IQ TS User Manual
Page 20
Commissioning
NitraVis 70x IQ TS
3 - 10
ba75950e02
11/2012
Cal - # raw value 1
Value pairs of the user calibration (see section 4.3.2).
Note: Leave those values unchanged during the initial commission-
ing.
Cal - ref. value 1
Cal - # raw value 2
Cal - ref. value 2
Zero adjustment
Factory
User
Setting of which zero adjustment is
used in the sensor.
Factory zero adjustment or
Zero adjustment carried out last
Signal smoothing
On
Off
Switching on/off the signal smoothing.
For more details, see
Response time
t90
.
Response time t90
10 ... 60 min
Response time of the signal smoothing.
Depending on the sample matrix, mea-
sured values can vary more or less
strongly (e.g. due to foreign bodies or
air bubbles).
The signal filter (here, the
Response
time t90
)
reduces the limits of variation
of the measured values.
The
Response time t90
specifies the
time after which 90 % of a signal
change are displayed.
This menu item is only available with:
Signal smoothing
=
On
The specified time is used as a guide
value for low (10 min) to high (60 min)
signal smoothing and can deviate from
the value that was set.
Cleaning mode
On
Off
Switching on/off the automatic, time-
controlled cleaning function.
Menu item
Settings
Explanations