NitraVis 70x IQ TS
Technical data
8 - 1
ba75950e02
11/2012
8
Technical data
8.1
Measurement characteristics
Measuring principle
Spectrophotometric absorption measurement; integrated micropro-
cessor electronics, shielded 2-wire connection for
power and data transmission.
Wavelength range
200 - 720 nm
Measuring gap (optical
layer thickness)
Nitrate measurement,
1 mm gap width
Nitrate measurement,
5 mm gap width
TSS measurement,
1 mm gap width
TSS measurement,
5 mm gap width
The quoted measuring ranges are nominal measuring ranges that are
theoretically possible. In practice, real measuring ranges exist that are
given by the limits of photometric determination. The limits are signifi-
NitraVis 701 IQ TS
1 mm
NitraVis 705 IQ TS
5 mm
Measuring
location
Measuring
mode
Measuring range
Resolution
Inlet
NO3
0.0 ... 300.0 mg/L
0.1 mg/L
NO3-N
0.00 ... 60.00 mg/L
0.01 mg/L
Activation
NO3
0.0 ... 300.0 mg/L
0.1 mg/L
NO3-N
0.00 ... 60.00 mg/L
0.01 mg/L
Outlet
NO3
0.0 ... 750.0 mg/L
0.1 mg/L
NO3-N
0.0 ... 150.0 mg/l
0.1 mg/L
Measuring
location
Measuring
mode
Measuring range
Resolution
Outlet
NO3
0.0 ... 250.0 mg/L
0.1 mg/L
NO3-N
0.00 ... 50.00 mg/L
0.01 mg/L
Measuring
location
Measuring
mode
Measuring range
Resolution
Inlet
TSS
0.00 ... 15.00 g/L
0.01 g/L
Activation
TSS
0.00 ... 20.00 g/L
0.01 g/L
Measuring
location
Measuring
mode
Measuring range
Resolution
Outlet
TSS
0.0 ... 900.0 mg/L
0.1 mg/L