Measurement
Fix the adapter for 16 mm vials on the sample chamber.
Turn	the	unit	on	using	the	[ON/OFF]	key.
The	display	shows	the	following:
Select	the	required	test	using	the	[MODE]	key.
scroll Memory (sM)
To	avoid	unnecessary	scrolling	for	the	required	test	range,	the	instrument	 
memorizes the last range
used before being turned off. When the instrument is
turned on again, the scroll list comes up with the last used test range.
The	display	shows	the	following:
Place	the	blank	in	the	adapter	(Note	1–4)	making	sure	that	the	 
marks are aligned.
Blanks are specially prepared for each individual test range.
Press	the	[ZERO/TEST]	key.
The	“Range“	symbol	flashes	for	approx.	8	seconds.
The	display	shows	the	following:
After	zeroing	remove	the	vial	from	the	adapter.
Place	the	sample	in	the	adapter	(Note	2–4)	making	sure	that	the	 
marks	 are aligned.
Press	the	[ZERO/TEST]	key.
The	“Range“	symbol	flashes	for	approx.	3	seconds.
the result appears in the display.
lr-/Mr-range:
in mg/l
Hr-range:
in g/l
the result is saved automatically.
Tolerance: ± 3.5 % (full scale)
repeating the test:
Press the [ZERO/TEST] key again.
new zero calibration:
Press the [ZERO/TEST] key for 2 seconds.